The initial analysis step detects the I-line interface rather
than the desired M-line interface, causing the analysis of
IMT to fail
Solution: Follow the Interactive Help
advice - during the 1st "setup" step when the
M-line is incorrectly detected, point to the proper location
along the M-line border and right-click. Once the M-line
border is correctly detected, proceed to the I-line detection
and sequence analysis.
-
Some AVI image data can not be opened.
Solution: Higher level of compression is not supported in order to achieve good quality of the analysis. Adjust the application that generates the AVI image data to use the non-compressed or minimal-compressed option for the image data.
|
|
|
|
|